Field of measurement: electricity and magnetism
Quantity: hf power ratio (attenuation, gain)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
9.26 (75 - 90) dB (0,12+0,019 (MH-75) - 0,12+0,019 (MH-75)) dB
  • frequency 18 GHz up to 26,5 GHz
113-MP-C014
9.26 (90 - 100) dB (0,12+0,012 (MH-90) - 0,12+0,012 (MH-90)) dB
  • frequency 0,1 MHz up to 13,2 GHz
113-MP-C014
9.26 (0 - 40) dB (0.1 - 0.1) dB
  • frequency 26,5 GHz up to 40 GHz
113-MP-C014
Quantity: antenna factor
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(-60 - 90) dB (0.14 - 1) dB
  • frequency 10 Hz to 30 MHz
113-MP-C016
Field of measurement: time and frequency
Quantity: frequency
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
1 (0.01 - 3.0E+9) Hz (1.0E-11 - 1.0E-11) 113-MP-C007
1 (3 - 18) GHz (1.0E-11 - 1.0E-11) 113-MP-C007
1 (18 - 46) GHz (3 - 3) Hz 113-MP-C007
Quantity: period
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
2 (5.0E-9 - 1.0E+5) s (1.0E-11 - 1.0E-11) 113-MP-C007
Quantity: time interval
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
3 (0 - 1.0E+5) s (1,1×10-9 + 1×10-11 t - 1,1×10-9 + 1×10-11 t) s 113-MP-C007
3 (0 - 10) s (10×10-12 + 2×10-3 t - 10×10-12 + 2×10-3 t ) s 113-MP-C007
Quantity: pulse counting
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
4 (0 - 1.0E+7) s 0
  • fmax=50 MHz 
113-MP-C007
Quantity: frequency deviation of frequency modulation
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
5 (0.2 - 40) kHz (1.5 - 1.5) %
  • fc=250 kHz up to 10 MHz, fmod: 20 Hz up to 10 kHz, Df/fmod ˃ 0,2 
113-MP-C007
5 (0.2 - 40) kHz (1 - 1) %
  • Df/fmod ˃ 1,2 
113-MP-C007
5 (0.25 - 400) kHz (1.5 - 1.5) %
  • fc=10 MHz up to 6,6 GHz, fmod: 50 Hz up to 200 kHz, Df/fmod ˃ 0,2 
113-MP-C007
5 (0.25 - 400) kHz (1 - 1) %
  • Df/fmod ˃ 0,45 
113-MP-C007
Field of measurement: metrology in chemistry
Quantity: Concentration
Internal identifier Range Calibration and measurement capabilities Analyt nebo komponenta Matrice Procedure Note to procedure
0.14 - 0.14 mg.l-1 0.006 - 0.006mg.l-1 ethanol in the gas phase 114-MP-C004
0.48 - 0.48 mg.l-1 0.011 - 0.011mg.l-1 ethanol in the gas phase 114-MP-C004
0.9 - 0.9 mg.l-1 0.019 - 0.019mg.l-1 ethanol in the gas phase 114-MP-C004
1.4 - 1.4 mg.l-1 0.028 - 0.028mg.l-1 ethanol in the gas phase 114-MP-C004
0.14 - 0.14 mg.l-1 0.008 - 0.008mg.l-1 ethanol in the gas phase 114-MP-C004
Field of measurement: ionising radiation
Quantity: Activity per unit area (C-14)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2001 (1 - 100) Bq.cm-2 3.6 %
  • C-14 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Co-60)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2002 (10 - 100) Bq.cm-2 3.2 %
  • Co-60 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Sr-90)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2003 (10 - 100) Bq.cm-2 3.2 %
  • Sr-90 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Cs-137)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2004 (10 - 100) Bq.cm-2 3.2 %
  • Cs-137 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Pm-147)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2005 (10 - 100) Bq.cm-2 3.2 %
  • Pm-147 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Tl-204)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2006 (10 - 100) Bq.cm-2 3.6 %
  • Tl-204 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Pu-239)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2007 (1 - 10) Bq.cm-2 3.6 %
  • Pu-239 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Am-241)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2008 (1 - 10) Bq.cm-2 3.6 %
  • Am-241 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (U-nat)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2009 (1 - 5) Bq.cm-2 4 %
  • U-nat rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Efficiency of contamination monitors (C-14)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2010 (0 - 1) s-1.Bq-1 10 %
  • C-14 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Cs-1137)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2011 (0 - 1) s-1.Bq-1 10 %
  • Cs-137 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Sr-90)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2012 (0 - 1) s-1.Bq-1 10 %
  • Sr-90 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Am-241)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2013 (0 - 1) s-1.Bq-1 10 %
  • Am-241 detection alpha radiation
115-MP-C007
Quantity: Activity per unit mass (H-3)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2014 (5000 - 5.0E+7) Bq.g-1 4 %
  • H-3 tritiated water
115-MP-C007
Quantity: Activity per unit mass (Be-7)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2015 (5000 - 5.0E+7) Bq.g-1 2 %
  • Be-7 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (C-14)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2016 (5000 - 5.0E+7) Bq.g-1 3 %
  • C-14 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Na-22)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2017 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Na-22 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Na-24)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2018 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Na-24 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (P-32)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2019 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • P-32 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (S-35)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2020 (5000 - 5.0E+7) Bq.g-1 2.6 %
  • S-35 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (K-42)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2021 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • K-42 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ca-45)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2022 (5000 - 5.0E+7) Bq.g-1 2.4 %
  • Ca-45 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cr-51)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2023 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Cr-51 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Mn-54)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2024 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Mn-54 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Fe-55)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2025 (5000 - 5.0E+7) Bq.g-1 5.5 %
  • Fe-55 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-56)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2026 (5000 - 5.0E+7) Bq.g-1 4 %
  • Co-56 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-57)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2027 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Co-57 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-58)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2028 (5000 - 5.0E+7) Bq.g-1 2 %
  • Co-58 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Fe-59)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2029 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Fe-59 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-60)
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EUR-RAD-CMI-2030 (5000 - 5.0E+7) Bq.g-1 0.8 %
  • Co-60 glass ampoule
115-MP-C007

Pages