You are here

CMC database

Field of measurement: mass
Quantity: mass of weights
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
1 (50 - 50) mg (0.0013 - 0.0013) mg 612-MP-C131
1 (100 - 100) mg (0.0016 - 0.0016) mg 612-MP-C131
1 (200 - 200) mg (0.002 - 0.002) mg 612-MP-C131
1 (500 - 500) mg (0.0026 - 0.0026) mg 612-MP-C131
1 (1000 - 1000) mg (0.003 - 0.003) mg 612-MP-C131
1 (2 - 2) g (0.004 - 0.004) mg 612-MP-C131
1 (5 - 5) g (0.005 - 0.005) mg 612-MP-C131
1 (10 - 10) g (0.006 - 0.006) mg 612-MP-C131
1 (20 - 20) g (0.008 - 0.008) mg 612-MP-C131
1 (50 - 50) g (0.01 - 0.01) mg 612-MP-C131
1 (100 - 100) g (0.016 - 0.016) mg 612-MP-C131
1 (200 - 200) g (0.03 - 0.03) mg 612-MP-C131
1 (500 - 500) g (0.08 - 0.08) mg 612-MP-C131
1 (1000 - 1000) g (0.15 - 0.15) mg 612-MP-C131
1 (2 - 2) kg (0.3 - 0.3) mg 612-MP-C131
1 (5 - 5) kg (0.8 - 0.8) mg 612-MP-C131
1 (10 - 10) kg (1.5 - 1.5) mg 612-MP-C131
1 (20 - 20) kg (3.3 - 3.3) mg 612-MP-C131
1 (50 - 50) kg (8 - 8) mg 612-MP-C131
Field of measurement: thermometry and humidity
Quantity: teplota
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
12 (-50 - 1760) °C 0.12 °C 611-MP-C130
12 (-50 - 1760) °C 0.16 °C 611-MP-C130
12 (0 - 1820) °C 0.14 °C 611-MP-C130
12 (-210 - 1200) °C 0.04 °C 611-MP-C130
12 (-270 - 400) °C 0.04 °C 611-MP-C130
12 (-270 - 1000) °C 0.04 °C 611-MP-C130
12 (-270 - 1370) °C 0.05 °C 611-MP-C130
12 (-270 - 1300) °C 0.05 °C 611-MP-C130
12 (-200 - 100) °C 0.03 °C 611-MP-C130
12 (-200 - 800) °C 0.03 °C 611-MP-C130
12 (0 - 2500) °C 0.24 °C 611-MP-C130
12 (-200 - 900) °C 0.04 °C 611-MP-C130
12 (-200 - 850) °C 0.02 °C 611-MP-C130
12 (-200 - 850) °C 0.02 °C 611-MP-C130
12 (-200 - 850) °C 0.03 °C 611-MP-C130
12 (-200 - 850) °C 0.02 °C 611-MP-C130
12 (-60 - 250) °C 0.01 °C 611-MP-C130
12 (-60 - 250) °C 0.01 °C 611-MP-C130
Field of measurement: electricity and magnetism
Quantity: DC voltage sources: single values
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
DCV01 (1 - 10) V (0,06 - 0,1) µV/V
  • fixed voltage 1 V, 1,018 V, 10 V
611-MP-C031
Quantity: DC voltage sources: low values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCV02 (0.001 - 10) V (0,051 - 8) µV 611-MP-C097
Quantity: DC voltage sources: intermediate values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCV03 (10 - 1100) V (0,8 - 1,6) µV/V 611-MP-C097
Quantity: DC voltage meters: very low values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCV05 (0 - 1) mV (100 - 100) nV 611-MP-C097
Quantity: DC voltage meters: intermediate values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCV06 (0,001 - 1100) V (0,1 - 2200) µV 611-MP-C097
Quantity: DC resistance standards and sources: multiple ranges
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCR05 (1 - 1.0E+8) W (1,5 - 55) µW/W 611-MP-C097
Quantity: DC current sources: intermediate values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCA02 (0.0001 - 20) A (6 - 15) µA/A 611-MP-C097
Quantity: DC current sources: high values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCA03 (20 - 100) A (20 - 20) µA/A 611-MP-C097
Quantity: DC current meters: intermediate values
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
DCA07 (0,0001 - 20) A (11 - 54) µA/A 611-MP-C098
Quantity: capacitance: low loss capacitors
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
CAP01a (1 - 1.0E+4) pF (0,5 - 20) µF/F
  • frequency 1 kHz
611-MP-C030
Quantity: capacitance: dissipation factor for low loss capacitors
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
CAP01b (1.0E-6 - 1) (3 - 300) 1.0E-6
  • capacitance 1 pF to 10 nF
  • frequency 1 kHz
611-MP-C030
Quantity: capacitance: dielectric capacitors
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
CAP02a (10 - 1.0E+4) nF 50 µF/F
  • frequency 1 kHz
611-MP-C030
Quantity: capacitance: dissipation factor for dielectric capacitors
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
CAP02b (1.0E-6 - 1) (40 - 400) 1.0E-6
  • capacitance 10 nF to 1000 nF
  • frequency 1 kHz
611-MP-C030

Pages